Calibration methods of new device for human blood pressure measurement
Wroclaw University of Science and Technology, Faculty of Mechanical and Power Engineering, Department of Criogenics, Aeronautics and Process Engineering, 50-370 Wroclaw, Wybrzeze Wyspianskiego 27, Poland
* Corresponding author: lukaszstala2@gmail.com
Published online: 4 February 2019
The paper presents two methods of determining calibration curve of a new device for blood pressure measurement. The device was developed at Wrocław University of Science and Technology. First method is based on parallel measurement of systolic and diastolic pressure measurement with use of reference device such as sphygmomanometer and researched new device with pneumatic sensor equipped with voltage type output. Obtained data (systolic ps and diastolic pd pressure, maximum us and minimum ud voltage) was then used to determine individual pressure-voltage characteristic of the device, which can be represented as a linear equation. Second method is based on substitution of experimentally proved coefficient b with its analytical equivalent extracted from mathematical model of described pneumatic sensor. Described methods were verified experimentally and compared. Metrological parameters of the device were designated.