High-count-rate 3He position-sensitive detector system: NEUNET-HCR
1 High Energy Accelerator Research Organization, 1-1 Oho Tsukuba Ibaraki, Japan
2 Canon Electron Tubes & Devices Co., Ltd., 1385 Shimoishigami Ohtawara Tochigi, Japan
3 CLEAR PULSE Co., Ltd., 6-25-17 Chuo Ota-ku Tokyo, Japan
* Corresponding author: setsuo.satoh@kek.jp
Published online: 11 March 2020
Highly reliable 3He gas detectors are most often used in neutron scattering experiments but have disadvantages such as low count rates and low position resolutions. In this study, a 3He gas detector system called NEUNET-HCR was developed, whose count rate is approximately 10 times that of the current NEUNET system, which was developed for a high position resolution with a low count rate. NEUNET-HCR achieved a maximum count rate of 535 kcps per detector, with a loss of approximately 30%. As the NEUNET-HCR system was developed based on the NEUNET system, various improvements were required to achieve the high count rate. Further, a high count rate was achieved in the NEUNET-HCR system without significantly degrading the position resolution.